Page 12 The Probe – INSTRUCTION MANUAL A5E32243983
mmmmm
English
Supplement
Measurement Interval
Troubleshooting
Patents
Instrument Housing Design:
Canada: 70345
U.S.A.: 07/858/707
Germany: M92022723
U.K.: 2021748
France: 921873
Japan: 966217
Electronics / Sensor:
U.S.A.: 5,267,219
5,339,292
U.K.: 2,260,059
patent applications in U.K., Canada, Europe, Africa, Australia
The echo is not reliable and The Probe is waiting for a valid echo before updating the
measurement.
Probable causes are:
material or object in contact with sensor face
The Probe is too close to the fill point
The Probe is not perpendicular to the liquid surface
change in level too fast
measurement out of range
foam on liquid surface
high level of vibration in the mounting structure
level inside the blanking zone
The `Waiting' period has expired. Investigate the probable causes listed above.
Refer to Speed of Response on page 10 or Fail-safe Timer on page 11 for duration of
`Waiting' periods.
interval
seconds
mA loop current