18. Substances which could damage the material of the centrifuge, the rotors or
the adapters must not be centrifuged. Infectious, toxic, pathogene or
radioactive substances must be centrifuged in certified rotors and vessels
only and all necessary safety precautions are taken.
19. Operation with incompletely loaded swing-out rotor or angle rotor with
interchangeable buckets.
A rotor must always be loaded completely, empty places are not allowed!
Opposite buckets or carriers may nevertheless be empty. Mixed loading is
allowed, if opposite places are loaded with same buckets and carriers of
same weight.