6 Malfunctions and Error Correction
6.1 Error Mode
Malfunctions are indicated by error messages with a number in the speed
display. If the sound signal is activated, it sounds when the error message is
displayed (see 5.3.8 "Activating/Deactivating the Sound Signal", page 38).
6.2 Error Correction
Eliminate the source of the problem (see 6.2.1 to 6.2.6, page 39ff).
Acknowledge the error messages with the lid key.
6.2.1 No Indication on the Display
Voltage in the socket? Check fuse in the mains supply.
Power cord plugged in and line
voltage present?
Plug in power cord correctly.
Fuse OK?
Replace fuse (see fig. 5.4 "name plate",
page 24).
Mains power switch on? Switch on power.
Lid closed? Close lid (see 6.2.4 “Lid cannot be
opened", page 40).
6.2.2 Centrifuge cannot be started
Start key LED not illuminated:
Power off/on. If the error occurs again,
call the service.
Lid key LED flashes:
Open and close the lid again. If error
occurs again although both locks are
locked, call the service.
6.2.3 Centrifuge decelerates during Operation
There was a brief mains power
failure (at least 2 sec), error
message 61
Centrifuge displays an error from
1 to 11 after powering on.
Press the start key in order to restart
the centrifuge.
Power off/on (see 6.2.6 "Error Codes",
page 41). If the error occurs again, call
the service.