Version 04/2008, Rev. 1.18 of 07/06/2012 page 41 of 65
Translation of the original operating manual
6.2.6 Error Codes
Error no.
Kind of error
Measure
Note
1-9 System error
•
allow to slow down
• power off/on
All these errors stop
the centrifuge or cause
it to slow down.
10-19 Speedometer error
•
allow to slow down
• power off/on
20-29 Motor error
•
power off
• ensure ventilation
30-39 EEPROM error
•
allow to slow down
• power off/on
With errors 34,35,36,
the centrifuge will stop.
With errors 37,38,
error message only.
40-45 Temperature error
(only for refrigerated
centrifuges)
•
allow to slow down
• power off
• allow to cool down
• provide better ventilation
46-49 Imbalance error
•
allow to slow down
• power off
• eliminate the imbalance
50-59 Lid error
•
press lid key
• close lid
• remove foreign matter
from the opening of the
lid lock device
With error 50 and 51,
the centrifuge will stop.
60-69 Process error
•
allow to slow down
• power off/on
With error 60,
message “power
failure during run”.
With error 61,
message “stop after
power on”.
70-79 Communication error
•
allow to slow down
• power off/on
80-99 Parameter error
•
power off
• allow to slow down
• provide for better
ventilation
With error 83, error
message only.
Attention!
If it is impossible to eliminate the errors, please call the service!
Fig.6.2: Error Codes