Operating Manual SIGMA 3-16K, page 19 of 69
03/06
Part No.
Description
Max. speed
(rpm)
Max.
gravitational
field (x g)
14002 Casset for 10 reaction vials 1.5 - 2.2 ml
15008, 15040, suitable for 11136,
polyallomer
14020 Casset for 10 PCR tubes 0.2 - 0.5 ml,
suitable for 11136, polypropylene
14021 Casset for 32 PCR tubes 0.2 ml or 4 strips
with 8 PCR tubes 0.2 ml each, suitable for
11136, polypropylene
11135 Swing-out rotor 24 x 1.5 - 2.2 ml for reaction
vials e.g. 15008, 15040, incl. hermetic
aluminium lid, max. radius 7.4 cm, min.
radius 3.5 cm
14 000
16 215
12154 Angle rotor 24 x 1.5 - 2.2 ml for reaction
vials e.g. 15008, 15040, incl. hermetic
aluminium lid, max. radius 8.2 cm, min.
radius 5 cm, angle 45°
15 300
21 460
12131 Angle rotor 30 x 1.5 - 2.2 ml for reaction
vials e.g. 15008, 15040, incl. hermetic
aluminium lid, max. radius10 cm, min.
radius 6.7 cm, angle 45°
14 000
21 913
12348 Angle rotor 48 x 1.5 - 2.2 ml for reaction
vials e.g. 15008, 15040, incl. hermetic
aluminium lid, max. radius 8.4 cm, min.
radius 5.1 cm
15 300
21 984
12115 Angle rotor for 12 strips with 8 PCR tubes
0.2 ml each, incl. hermetic aluminium lid,
max. radius 9.8 cm, min. radius 7.2 cm,
angle 45°
14 000
15 777/21 475
12111 Angle rotor 10 x 10 - 12 ml for sealed tubes
15000, 15010, 15039, incl. hermetic
aluminium lid, max. radius 7.6 cm, min.
radius 2.8 cm, angle 35°
15 300
19 890
12157 Angle rotor 20 x 10 - 12 ml for sealed tubes
15000, 15010, 15039, incl. hermetic
aluminium lid, max. radius 9.8 cm, min.
radius 5.9 cm, angle 25°
10 000
10 956