Operating Manual SIGMA 3-16K, page 62 of 69
03/06
7.2 Error Correction
Most of the errors can be reset by power off/on. In the event of a short power
failure during a run, this run is interrupted and can be continued by pressing the
Start-key.
No indication on the display: Actions:
No power in the socket?
Power cord plugged in and line
voltage present?
Power switch on?
Lid closed?
Check fuse in mains supply.
Plug in power cord correctly.
Switch on power.
Close lid (please refer to chapter 7.2.3 „Lid
cannot be opened/closed.“).
7.2.1 Centrifuge cannot be Started
Start-key LED illuminated: Power off/on. If error occurs again, call
service.
Lid-key LED flashing: Open and close the lid again. If error
occurs again although the lock engaged,
call service.
7.2.2 Centrifuge Decelerates during Operation
Centrifuge displays an error 1
to 11 after power on.
Power off/on. If error occurs again, call
service (please refer to chapter 7.3.1
"Error codes").
7.2.3 Lid cannot be Opened/Closed
When first trying to open the lid the lock is not released. The Lid-key LED is
flashing. Open and close lid again.