7.2.4 Emergency Lid Release
At the upper right panel there is a stopper which can be removed eg. with a
screw driver. The lid can be released by pulling the visible string.
Attention!
The lid may only be unlocked and opened when the rotor is at standstill.
7.2.5 Problems with the Centrifuge?
Please contact your supplier for support or in the event of malfunctions and for
supply of spare parts.
7.3 Error Mode
In the error mode the Start-, Stop- and Lid-keys are flashing. “ERROR“ is
displayed in the speed area and the error code in the rcf area.
7.3.1 Error Codes
Error no. Kind of error Actions
1 Tacho signal disturbed
Power off/on
4 Lid does not open after
pressing the Lid-key
Power off/on
Emergency lid release
2, 3, 5 – 11, 17 - 30 Internal fault
Power off/on
12 Imbalance determined
Remove imbalance
13 Temperature sensor
Contact service
14 Excess temperature (>50°C)
Let centrifuge cool down
Should it not be possible to repair the failure, please contact Service!
7.4 Speed-RCF-Diagram