3.3 Installation of Rotors and Accessories
1. Open centrifuge lid by pressing Lid-key.
2. Unscrew rotor tie-down screw from motor shaft (anticlockwise).
3. Lower the rotor straight down onto the motor shaft.
4. Tighten the tie-down screw (clockwise) with the rotor wrench so that the disc
spring is pressed together.
Fastening torque: 1-6/1-13/1-15/2-5/1-15K/2-16/2-16K/
2-16KC:
approx. 5 Nm
3-16/3-16K/3-18K/3K 30: approx. 7,5 Nm
4-15/4K15/6-15/6K15: approx. 10 Nm
In the event of frequent use the tie-down screw must be loosened by some
turns and fastened again. This should be done once a day or after approx.
20 cycles. This ensures a proper connection between rotor and shaft (please
refer to chapter 6.2 "Care and cleaning of accessories" as well).
5. Fill all positions of swing-out
rotors with buckets. Make
sure that all buckets are
inserted correctly.
6. Use only appropriate vessels
for the rotor (please refer to
chapter 1.2 "Suitable acces-
sories" as well).
Universal grease
allowed not allowed
7. Fill vessels external to the centrifuge.
8. Put or screw on covers of vessels.
9. Opposite places of the rotors must always be loaded with same accessories
and same filling.
10. In angle rotors the plastic vessels must always be totally filled to avoid cracks
of vessels and leakages or loosening of the caps in case of partial filling.
Attention, follow the special comments of chapter 1.5.
11. Attention: The centrifuge will absorb smaller differences in weight when
loading the rotors. But it is recommended to balance the vessels as
accurately as possible in order to ensure a run with minimal vibrations.
Should the centrifuge be operated with very uneven load, the imbalance
monitoring will switch off the drive. An imbalance warning would be displayed
(refer to 4.1).