Page 42
Wiring the Sampler
Section 2
Figure 17 Connecting to a Rain Gauge
2.12 Wiring the Sampler
1. Disconnect all power to the 980 Flow Meter. See Wiring Safety
Information on page 31.
2. Use a large flat-blade screwdriver to loosen the two screws securing the
980 Flow Meter cover. Open the cover.
3. Strip insulation from the sampler leads ¼ inch.
Note: Route wires through
NEMA-approved conduit hubs
(Cat. No. 16483) to ensure
that water and dust do not enter the
enclosure.
4. Attach a NEMA-approved conduit or compression fitting to one of the
½ in. openings on the bottom of the instrument, and route the wires
through this opening.
5. Connect wires to the proper screw terminal block (TB10).
Refer to Ta bl e 8 and Figure 18 for wire connections.
Cable Required for Sampler Connections
Multi-Purpose Half Cable Assembly, 10 ft (3.0 m), 6-pin connector on one
end, tinned wire leads on the other end (Cat. No. 9708700).
1. Strain Relief 2. Input from Rain Gauge
J12
J13
TB8
XMIT+
CH A+CH 4
RAIN+
1
1
1
XMIT–
RCV+
RCV–
SHEILD
AVTB11
TB10 4-20 mA
TB 9 ANALOG
TB4
RAIN
COMMON
CH A–GROUNDRAIN–
+12V
SHIELDSHIELDSHIELD
DEPTH–
CH B+CH 5CH 1
DEPTH+
CH B–GROUNDGROUND
SHIELD
SHIELDSHIELDSHIELD
COMMON
STARTCH 6CH 2
DSR
PULSEGROUNDGROUND
RXD
BOTTLESHIELDSHIELD
SHIELD
SHIELD
TOTALIZER
TB7
TB6
TB2
TB5
TOTAL –
RTD –
TIP
TIP
TOTAL +
1
1
MODEM
ULTRASONIC
RTD
RING
RING
RTD
DTR
EVENTCH 7CH 3
TXD
GROUNDGROUNDGROUND
+12V
SHIELDSHIELDSHIELD
RS232C
SAMPLERANALOGANALOG
1
1
1
1
2
2
2
2
3
3
3
3
4
4
4
4
5
5
5
5
6
6
6
6
7
7
7
7
8
8
8
8
9
9
9
9
10
10
10
10
11
11
11
11
12
12
12
12
13 14 15 16
SHIELD
TB3
pH
REF
pH
RTD
–5V
GROUND
+5V
485A
485B
485+NO
NO
NO
NO
485GNDCOM RELAY 1
COM RELAY 2
COM RELAY 3
COM RELAY 4
SHIELDNC
NC
NC
NC
RS485
12345 6 789101112
F2
F1
HOT/ 1
NEU/ 2
T, 1.0A 250V
ON
HOT/ 1
NEU/ 2
TB8
1
4
7
10
2
5
8
11
3
6
9
12
J13
RAIN+
1
TB4
RAIN
RAIN–
SHIELD
CH 1
GROUND
SHIELD
CH 2
GROUND
SHIELD
CH 3
GROUND
SHIELD
ANALOG
12345 6 789101112
POWER
OFF
2
1
No Connection

Table of Contents