Operating
Instructions
D.C.
VOLTAGE
MEASUREMENTS,
0
1000
VOLTS
18
14
Set
the
function
switch
оп
the
left
hand
side
of
the
front
panel
at
+D.C.
Plug
the
black
test
lead
into
the
COMMON
-
jack
and
the
red
test
lead
into
the
*
Jack.
Set
the
range
selector
switch
in
any
of
the
five
voltage
range
positions.
These
are
marked
2.5
V.,
10
V.,
50
V.,
250
V.,
and
1000
V.
When
in
doubt
as
to
the
voltage
present,
always
use
the
highest voltage
range
as
a
protection
to
the
instrument.
Observe
the
meter
reading.
If
the
voltage
is
within
a
lower
range,
the
switch
may
be
set
for
the
lower
range
to
obtain
a
more
accurate
reading.
Connect
the
black
test
lead
to
the
negative
side
of
the
circuit
to
be
measured
and
the
red
test
lead
to
the
positive
side
of
the
circuit.
Turn
on
the
power
in
the
circuit
to
be
tested.
If
the
pointer
deflects
to
the
left
of
zero,
the
anticipated
polarity
is
opposite
the
actual
circuit
polarity.
Turn
off
the
power
in
the
circuit
which
is
being
tested.
Set
the
function
switch
at
D.
C.,
and
turn
on
the
power
again.
This
will
correct
the
polarity
as
applied
to
the
meter.
Read
the
voltage
on
the
black
arc
marked
D.C.
which
is
second
from
the
top
of
the
dial.
For
the
2.5
V.
range,
use
the
0—250
figures
and
divide
by
100.
For
the
10
V.,
50
V.,
and
250
V.
ranges,
read
the
figures
directly
on
the
scale.
For
the
1000
V.
range,
use
the
0—10
figures
and
multiply
by
100.