Operating
Instructions
ADJUST
POINTER
FOR
ZERO
Before
any
measurements
are
made,
check
to
see
that
the
pointer
indicates
zero
when
the
meter
is
in
its
operating
position.
If
the
pointer
is
off
zero,
adjust
the
screw
located
in
the
phenolic
case
below
the
center
of
the
meter
scale,
as
shown
in
Figure
1.
Use
a
small
screwdriver
to
turn
this
screw
slowly
clockwise
or
counterclockwise
until
the
pointer
is
exactly
over
the
zero
mark
at
the
left
side
of
the
scale.
D.C.
VOLTAGE
MEASUREMENTS,
0-250
MILLIVOLT
RANGE
ONLY
.
Set
the
function
switch
at
+D.C.
Connect
the
black
test
lead
in
the
COMMON
-
jack
and
the
red
test
lead
in
the
SO&AMPS
jack.
.
Set
the
range
switch
at
50
HhAMPS
(common
position
with
50
V.).
Connect
the
black
test
lead
to
the
negative
side
of
the
circuit
to
be
measured,
and
the red
test
lead
to
the
positive
side
of
the
circuit.
.
Turn
on
power.
If
meter
deflects
to
left
of
zero,
turn
off
power
and
reverse
the
test
lead
connections.
Then
turn
on
power
again.
.
Read
the
voltage
on
the
black
arc
marked
D.C.,
and
use
the
figures
marked
0—
250;
read
directly
in
millivolts.
Turn
off
power
in
the
circuit
which
is
being
measured
before
disconnecting
meter
leads.