BS Series Multitest Analyzer
5
The default values of the instrument is that 3000ul aspiration sample volume corresponds
to 20000 motor steps. When the aspiration volume is not correct, adjusting is necessary.
Enter into the calibrate interface, input pump calibrate volume, then pour corresponding
distilled water into the tube, insert pipette into water, press PUSH to aspirate. After the
distilled water was sucked done. Press PUSH again, the instrument will display motor step
and exit pump calibration to save step. If the aspiration volume is still not right, repeat the
above operation or input moter steps directly.
3.4 Instrument operation key process
3.4.1 Instrument preheating
Connect to the power and turn on the switch, the instrument should be preheated in
10-30 mins.
3.4.2 Pipeline Washing
Clean the flow cell before using it, enter into the main interface, insert the pipette into
distilled water, and then press wash to start washing, last 5-10 times.
Attention:
Pay attention to use environmental dustproof, moisture proof, and installation air
conditioning is better, environment temperature in 18-25 ° C is the best working
temperature.
● When external power is not stable, instruments must connect regulated power supply
Instruments in the process of operation shall not open the cover, so as not to cause
damage to equipment or operator
Leakage and electrostatic prevention, the instrument should be in good grounding. Line
power socket must have a reliable grounding line to guarantee in a steady state and
security
After work, wash the instrument 3 times at least immediately to keep the cuvette and
pipeline from liquid waste.
After finish the testing, the used should be collected and disposed according to the
medical waste.
The waste pipe end should not be dipped into the waste to avoid poor drainage.
Use qualified reagent within the period of validity.