Remove the adhesive backing panel from the
RENASYS So Port dressing, and align directly
over the hole in the transparent lm.
Use gentle pressure to anchor the So Port
tothe transparent lm.
8
Secure the RENASYS So Port to the patient according
toyour institutional protocol.
Ensure the aeration disc, located near the orange quick
click connector, is not covered or otherwise occluded
bythemethod used to secure the So Port.
10
Connect the RENASYS So Port to the canister tubing
bypushing the orange quick click connectors together.
An audible click indicates connection is secure.
Activate the RENASYS pump ensuring that it is operating
at the prescribed therapy level. Please refer to your
RENASYS pump manual for full operating instructions.
Thenished dressing should be fully compressed,
rmtothetouch and leak-free.
11
Smooth the dressing down while removing
the RENASYS So Port stabilisation frame.
9
RENASYS
◊
EDGE Quick reference guide 6