8
ENG
SYSTEM SPECIFICATIONS
CAUTION – Only VERSAJET™ approved equipment should be connected to this device.
CONSOLE
Front Panel: Illuminated Power Switch, On/Off (1/0)
Foot Switch Receptacle
Speed Control for levels 1 through 10
Amber “Door” indicator Light
Amber “Pedal” indicator Light
Red “Error” indicator Light
Rear Panel: Power Cord Receptacle
Power Console: AC Power: Detachable power cord with a Hospital Grade connector
100-120 V / 200-240 V ~ 6A / 3A 50/60 Hz
Prior to connecting the device to an electrical mains outlet, determine local voltage
and electrical supply requirements. Ensure the cable is compatible.
Depress voltage switch toward direction indicated, 115V (100-120) or 230V (200-240)
Size: 18” W x 13” D x 8” H (45.7 cm W x 33.0 cm D x 20.3 cm H)
Weight: 28 pounds (12.7 kg)
FOOT SWITCH
Size: 3” W x 9” D x 2.5” H (7.6 cm W x 22.9 cm D x 6.4 cm H)
Weight: 3 pounds (1.12 kg)
POWER CORD
Length: 15 feet (4.5 m)
DISPOSABLE HANDPIECE ENVIRONMENTAL CONDITIONS
Unless otherwise stated, the following conditions apply for product use as well as shipping and handling:
Temperature Range (Shipping & Handling): -40°F (-40°C) to 125°F (52°C)
Temperature Range (Product Use): 40°F (4°C) to 100°F (38°C)
Humidity Range: 0% to 100%, noncondensing
Atmospheric Pressure: 500 to 1500 millibar
POWER CONSOLE ENVIRONMENTAL CONDITIONS
Unless otherwise stated, the following conditions apply for product use as well as shipping and handling:
Temperature Range (Shipping & Handling): -4°F (-20°C) to 131° F (55°C)
Temperature Range (Product Use): 40°F (4°C) to 100°F (38°C)
Humidity Range: 0% to 100%, noncondensing
Atmospheric Pressure: 500 to 1500 millibar
ELECTROMAGNETIC INTERFERENCE RISK
The power console has been tested according to IEC 60601-1-2 (2001-09)
115
230