acapella
®
choice Vibratory PEP System
Reference Guide
These instructions contain important information
for safe use of the product. Read the entire contents
of the Instructions for Use, including Warnings and
Cautions, before using this product. Failure to properly
follow warnings, cautions and instructions could
result in death or serious injury to the patient. It is the
responsibility of the healthcare practitioner to assure
that the instructions for use and maintenance are
understood by and provided to the caregiver.
B.
A.
D.
C.
Figure 1
Figure 2
Figure 3
Figure 4
Figure 5
90°
Rib
Figure 6
1. DESCRIPTION:
The acapella
®
choice vibratory PEP system is a
single patient use device that provides Positive
Expiratory Pressure (PEP) therapy for patients
who have Cystic Fibrosis, COPD, asthma, and lung
diseases with secretory problems, and patients
with atelectasis. All patients must be capable of
following instructions for PEP therapy. Review the
diagram and the product to become familiar with
all of the product features.
The acapella
®
choice Vibratory PEP system with
one-way inspiratory valve and 22 mm male fittings
consists of: (See Figure 1)
A. Detachable mouthpiece
(device end adapts to mask)
B. Expiratory resistance/frequency adjustment dial
C. Detachable cover
D. Detachable rocker assembly – DO NOT
disassemble the rocker arm from its
platform base
2. INDICATIONS:
The Smiths Medical acapella
®
choice system is
intended for use as a PEP device. It may also be
used simultaneously with nebulized aerosol
drug delivery.
3. CONTRAINDICATIONS:
Although no absolute contraindications to the use
of PEP therapy have been reported, the following
should be carefully evaluated before a decision is
made to initiate therapy:
Inability to tolerate increased work of breathing
Hemodynamic instability
Intracranial pressure (ICP) > 20 mm Hg
Acute sinusitis
Recent facial, oral or skull surgery or trauma
Epistaxis
Esophageal surgery
Active hemoptysis
Untreated pneumothorax
Nausea
Known or suspected tympanic membrane rupture
or other middle ear pathology
4. WARNINGS:
4.1 Use of this device at excessive pressures may
have adverse effects. Expiratory pressures above
20 cm H2O in patients sensitive to increased
transpulmonary pressure may develop one or
more of the adverse side effects listed below.
4.2 Expert clinical judgment should be exercised in
the selection of the appropriate setting for each
individual patient. Failure to match the appropriate
resistance setting on the +/- dial indicator with
the patient’s expiratory flow may result in failure
to achieve therapeutic objectives of vibratory PEP
therapy or one or more adverse side effects below.
Adverse reactions may include:
Increased work of breathing that may lead to
hypoventilation and hypercarbia
Increased cranial pressure
Cardiovascular compromise
Myocardial ischemia
Decreased venous return
Air swallowing with increased likelihood of
vomiting and aspiration
Claustrophobia
Skin break down and discomfort from mask
Pulmonary barotraumas
4.3 The rocker assembly is not designed to be
disassembled - do not attempt to separate
the rocker arm from its platform base. Separating
these components may cause the acapella
®
choice system to function improperly, which may
deprive the patient of therapy and cause injury to
the patient.
5. PRECAUTIONS:
5.1 Bleach is not recommended for use on the
acapella
®
choice system. It may deteriorate the
nickel plated mechanism located in the interior
of the device.
5.2 DO NOT MICROWAVE. The metal and magnet
might ignite.
5.3 It is the responsibility of the user to ensure all
sterility verification(s).
5.4 Visually inspect the device to ensure that the
unit is free of contamination and foreign objects.
5.5 Verify all connections are secure.
Question and Answer IconNeed help?

Do you have a question about the Smiths Medical acapella and is the answer not in the manual?

Smiths Medical acapella Specifications

General IconGeneral
BrandSmiths Medical
Modelacapella
CategoryMedical Equipment
LanguageEnglish

Summary

Product Description and Indications

Contraindications

Warnings and Precautions

Excessive Pressure Effects

Warning about adverse effects from excessive expiratory pressures above 20 cm H2O.

Clinical Judgment and Adverse Reactions

Importance of expert judgment in setting resistance and potential adverse reactions.

Rocker Assembly Warning

Warning against disassembling the rocker arm to prevent malfunction or injury.

Bleach and Microwave Use

Precautions against using bleach and microwave due to potential damage or ignition.

User Responsibility for Sterility and Inspection

User's responsibility for sterility verification and inspecting the device for contamination.

Connection Verification

Ensure all connections are secure for proper device function and safety.

Suggested Instructions for Use

Initial Device Settings

Guidance on setting the frequency adjustment dial for the first use of the device.

User Procedure and Breathing Technique

Step-by-step instructions for patient use, including breathing and exhalation techniques.

Nebulizer Set-up Integration

Instructions for integrating the acapella choice system with a nebulizer.

Cleaning and Disinfecting Instructions

Disassembly for Cleaning

Steps for safely disassembling the device components for cleaning purposes.

Cleaning and Disinfecting Methods

Guidelines for cleaning with detergent and disinfecting via boiling, autoclaving, or dishwasher.

Disposal and Device Lifetime

Proper Disposal Procedures

Instructions for safe disposal according to regulations for contaminated medical waste.

Estimated Device Lifetime

Information on the expected useful life of the device under normal usage.