THERMOCUBE 200/300/400 THERMOELECTRIC CHILLER MANUAL 52-11300-1
SOLID STATE COOLING SYSTEMS, 167 MYERS CORNERS ROAD, WAPPINGERS FALLS, NY 12590 10
TELEPHONE: (845) 296-1300 FAX: (845) 296-1303 WEB: WWW.SSCOOLING.COM VERSION M28
3.3 AIR CONSIDERATIONS
The air inlet and outlet are located on the left and right sides
respectively. Restricting airflow into or out of the unit will impair
performance. At least 3 inches of clearance is required on each side
to ensure adequate airflow.
An optional Air Filter is also available (see Section 7 for options).
3.4 COOLANT FILL
The coolant fill cap is located at the top rear of the unit. Twist off
the cap counter-clockwise to open. Fill reservoir prior to starting
unit. Close cap before operating
Recommended Coolants:
Solid State Cooling Systems recommends using Koolance, a pre-
mixed 27% propylene glycol/water based coolant containing an
algaecide and corrosion inhibitors. Though it comes in several
colors, SSCS recommends the colorless or blue versions in 700 ml
bottles, part number: LIQ-702CL-B (clear) or LIQ-702B-B (blue),
as the dyes in the other colored versions can form small particulates
when not well mixed.
Contact Koolance for details:
Koolance USA
2840 West Valley Highway North
Auburn, WA 98001
(253) 893-7551
Water or ethylene glycol/water mixtures are also acceptable as
coolants.
Note that algae growth can occur when water is used without at
least 25% propylene or ethylene glycol.
Prior to filling the chiller
Use only recommended
coolants