14 Probes and Biopsy
138 Basic User Manual
7. Press the handle and insert the biopsy bracket tube into the biopsy bracket.
8. Insert the biopsy needle into the guide tube and make sure that the biopsy bracket is firmly attached to the
probe.
9. Unfold another probe sheath, and apply an appropriate amount of coupling gel to the inside of the sheath.
10. Hold the probe attached with the biopsy bracket and unroll the sheath onto both the probe and attached biopsy
bracket.
■ To Assemble the Endocavitary Biopsy Bracket
Top Housing
Inner Part Guide Tube
Perform the following steps.
1. Check and ensure that there is no damage, deformation, malfunction, loose or missing parts on the biopsy
bracket before assembly.
2. Wear sterile gloves.
3. Unfold the probe sheath, and apply an appropriate amount of coupling gel into the sheath.
4. Hold the probe and unroll the sheath onto it. Pull the probe sheath tightly over the face of the probe to remove
any wrinkles.
5. Assemble the metal top housing and the biopsy bracket. Slide the inner part into the top housing and push
forward until it clicks.