Operation
DocNr.: 03-4655 R03.01 | 22.09.2015 | en-US | Ultrasonic Generator 31
Starting treatment
Contaminated instruments
The use of contaminated instruments may result in life-threatening infections in both
patients and users.
Use only a sterilized instrument.
Only use a sterilized instrument cable.
Use only sterilized accessories (irrigation tubing, cover).
Instrument remains active
An instrument that remains active because, for example, the foot switch does not
switch off may result in severe injury to the patient.
Hold the instrument away from the patient.
Turn off the ultrasonic generator.
Unintentional activation of the foot switch
An unintentionally activated instrument can result in severe injury to the patient.
Position the foot switch so it cannot be activated unintentionally or mistaken for
other foot switches.
Place the instrument on a utility table, not on the patient.