Span Medical Retractable Bed
MC9R / Series
27
Guidance and Manufacturer’s Declaration – Immunity
The MC9R/Encore
®
is intended for use in the electromagnetic environment specified below. The
customer or user of the MC9R/Encore
®
should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment -
Guidance
ESD
IEC 61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
Floors should be
wood, concrete or
ceramic tile. If floors
are synthetic, the r/h
should be at least
30%.
EFT
IEC 61000-4-4
±2kV Mains
±1kV I/Os
±2kV Mains
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1kV Differential
±2kV Common
±1kV Differential
±2kV Common
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage Dips/Dropout
IEC 61000-4-11
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
MC9R/ENCORE
®
requires
continued operation
during power mains
interruptions, it is
recommended that the
MC9R/ENCORE
®
be
powered
from an uninterruptible
power supply or
battery.
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
3A/m
Not Applicable
Power frequency
magnetic fields should
be that of a typical
commercial or hospital
environment.