Spectracom Corporation PathAlign-R
2200 / 2240 / 2241 PathAlign-R Manual
3.3 Rear Panel
Figure 3-2 depicts the PathAlign-R rear panel connectors and battery across door.
Figure 3-2. Model 2240 Rear Panel
Hinged battery access door provides for removal/replacement
of internal battery. Door latches via a thumbscrew.
Osc. Calib. Access hole allowing the use of a trim pot adjusting tool to fine-
tune the internal oscillator during yearly calibration. Note:
Refer oscillator calibration to a qualified technician!
Type 5.5 mm x 2.1 mm, two-terminal battery power connector
connects the external 15VDC/ 1A Battery Charger to the
PathAlign-R (center pin is positive, maximum char
BNC type connector. Provides a 0-2 VDC output that tracks the
front panel LCD Path Loss Meter (1 mV = 0.1 dB). Allows for
external readout of path loss using a handheld DVM.
3.4 Operating Characteristics
The following paragraphs describe the operating range, band resolution, dynamic range, and
overload characteristics.
3.4.1 Operating Ranges
There are seven operating ranges available: Band 1: 1.5 GHz to 2.5 GHz; Band 2: 3.5 GHz to
5.0 GHz and 5.8 GHz to 6.6 GHz; Band 3: 7.5 GHz to 10.0 GHz and 11.0 to 12.0 GHz; Band 4: