5.Press [6] to select Standard Curve menu item. Press [Enter].The LCD will display as blow.
6.Press [1] to select Create Curve menu item. Press [Enter].The LCD will display as blow.
7.Press [1] to select No. of samples menu item. Press [Enter].The LCD will display as blow.
Input
Sample Num:
1.No. of Samples
2.Unit
3.Input Conc
4.Test Samp le
1.Create Curve
2.Load Curve
3.Delete Curve
4.Print Curve
5.Lamp on/off
6.Standard Curve
7.Language
8.Com. With PC
8. Use digital key to enter samples number, press [Enter].
9. Press [2] to select Unit menu item. Press [Enter].The LCD will display as blow.
10.
Use digital key to enter sample unit, press [Enter].
11. Press [3] to select Input Conc menu item. Press [Enter].The LCD will display as blow.
Input
Conc:
Number:1
1.mg/L
2.ug/L
3.ng/kg
4.ug/kg
12. Use digital key to enter sample concentration, press [Enter].
13. Press [4] to select Test Sample menu item. Press [Enter].The LCD will display as blow.
Number: 1
Conc: 1.00
Abs.:
14.Insert the standard solutions in the light beam. Press [Enter].
User’s Manual SP-UV 200 22
Spectrum Instruments