23
Storing measurement results
Using the Save function to store measurement results is useful
for easy comparison of readings from different points in time,
thus making it possible to see the development of the condi-
tion for individual measuring points. VibChecker can store up
to ten vibration measurement results along with date/time,
ISO settings and spectrum.
Saving a reading
To save a reading after completed measurement:
1. In the Measurement display (A), press the LEFT arrow key
to enter the Save function (B).
2. Use the arrow keys to select a memory location (1-10).
3. Press the ENTER key to store the reading. This action will
overwrite any previously stored value in the memory ad-
dress selected. A checkmark symbol (C) confirms the save
operation, and the current memory location turns the same
color as the condition evaluation of the reading just saved.
4. To return to the Measurement display, press the BACK key.
The stored measurement results (excluding spectrum) can be
transferred from VibChecker to a Microsoft Excel file using a
downloadable SPM program (http://downloads.spminstru-
ment.com). The data in the Excel file can be imported into a
CMMS or other software for follow-up. For further instructions,
please see information in the delivery package.
As an alternative, the final pages of this User Guide contains
a follow-up form which can be copied and used to manually
record measurement results over time.
Viewing a stored reading
To view a stored reading:
1. In the Main display, select the View stored readings icon
(D) and press ENTER.
2. In the View stored readings display (E), use the arrow keys
to highlight a memory location.
3. Press the ENTER key to view the stored reading (F).
To return to the View stored readings display, press the
BACK key.
Deleting stored readings
To delete the stored readings:
1. In the Main display, select the View stored readings icon
(D) and press ENTER.
2. In the View stored readings display (E), use the arrow keys
to highlight the trash can (G).
3. Press the ENTER key to delete all of the stored readings.
To return to the View stored readings display, press the
BACK key.