MAX 4000 PLUS ELECTROMETER 2
General Precautions
Warnings and Cautions alert users to dangerous conditions that can occur if instructions in
the manual are not obeyed. Warnings are conditions that can cause injury to the operator,
while Cautions can cause damage to the equipment.
WARNING: Proper use of this device depends on careful reading of all instructions
and labels.
WARNING: An electrical shock hazard of up to 450 VDC is possible whenever the
bias voltage is active. Always set the bias to 0 VDC whenever a device is connected
to or disconnected from the MAX 4000 Plus.
WARNING: Electric shock hazard. Refer servicing to a qualified individual.
WARNING: The MAX 4000 Plus is not intended to be used in flammable mixture
atmospheres. Do not use with flammable anesthetic mixture with air, with oxygen,
or nitrous oxide.
WARNING: Where applicable, Standard Imaging products are designed to be used
with the versions of common radiation delivery devices, treatment planning
systems and other products or systems used in the delivery of ionizing radiation,
available at the time the Standard Imaging product is released. Standard Imaging
does not assume responsibility, liability and/or warrant against, problems with the
use, reliability, safety or effectiveness that arise due to the evolution, updates or
changes to these products or systems in the future. It is the responsibility of the
customer or user to determine if the Standard Imaging product can be properly
used with these products or systems.
CAUTION: Input voltage on triax connector should be no greater than +5 V or -5 V.
CAUTION: Only connect the RS-232 serial port to equipment where no risk of
external voltage exists.
CAUTION: This device should never be submerged in water or a solvent to clean, or
scrubbed with an abrasive cleaner. Do not drop or mishandle unit. Calibration factor
changes may result.
CAUTION: Do not disassemble unit since it may result in change of calibration
factor. Refer servicing to qualified individuals.
CAUTION: Damaged or kinked ionization chamber cables or extension cables
should not be used.
CAUTION: Upon power on, allow for proper warm up.
CAUTION: Charge measurements made in modes other than those used by the
calibration agency may result in measurement errors which exceed the calibration
uncertainty.
CAUTION: When connecting or disconnecting an ionization chamber to the MAX
4000 Plus, set the bias voltage to 0 VDC.
CAUTION: This user manual refers only to MAX 4000 Plus units. Refer to reference
number 80206 for MAX 4000 units.
CAUTION: For best performance, select the lowest range possible for the intended
measurement.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Standard Imaging MAX 4000 PLUS and is the answer not in the manual?

Standard Imaging MAX 4000 PLUS Specifications

General IconGeneral
BrandStandard Imaging
ModelMAX 4000 PLUS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals