6. Press under thickness-measuring state to enter into 2-Point
CAL, the screen will prompt to
calibrate the thinner piece.
7. Measure the thinner standard test piece, use or to
adjust the measurement to standard value. Press ENTER, the
screen prompts to measure the thicker piece.
8. Measure the thicker standard test piece, use or to adjust
the measurement to standard value.
9. Press , the calibration operation is finished.
6) Auto Down : Power-saving mode ON(default)
7) Baud Rate: 1200;2400;4800;9600
8) Adjusting Display Brightness
1. Press select “Set brightness”:
2. Use or to adjust the display brightness.
3. Press exit setup.
3.8.2 Print function
Connect main unit with micro printer from by the communication
cable, print measured results through menu selection.
When the printing is completed, the buzzer will give out sound, the
display will return to MENU state.
Send data to PC:
Connect main unit with PC by using the communication cable, one can