What to do if the Steris RELIANCE 400 Washer treatment stops for no apparent reason?
- AAlexandra GrantSep 1, 2025
If the treatment stops unexpectedly, check that the chamber doors are fully closed and close them securely.
What to do if the Steris RELIANCE 400 Washer treatment stops for no apparent reason?
If the treatment stops unexpectedly, check that the chamber doors are fully closed and close them securely.
How to resolve insufficient water entering Steris RELIANCE 400 Washer chamber through spray arms?
To resolve insufficient water entering the chamber through the spray arms, first ensure each accessory header is positioned directly over a manifold connector. Additionally, check the rotary spray arms and chamber debris screens for clogs and clean them if necessary.
How to resolve water leaks from the Steris RELIANCE 400 unit?
To address water leaks from the unit, tighten all clamps and connections.
What to do if the Steris RELIANCE 400 cycle does not start?
If the cycle does not start when you press the CYCLE/START touch pad, or if the pump doesn't start during treatment, ensure that the chamber door is closed. If the door was open, close it securely and then press the CYCLE/START touch pad again.
Why is there foam in the Steris RELIANCE 400 Washer chamber?
If you observe foam in the chamber, it might be caused by using the wrong type of detergent. Refer to the detergent specifications and follow the detergent manufacturer's recommendations for the amount of detergent to be used. Additionally, too much chemical injected during treatment can cause foam. Verify the chemical injection time in Automatic mode.
What to do if the Steris RELIANCE 400 Washer chamber does not drain completely?
If the chamber isn't draining completely, start by checking the chamber debris screens for clogs and clean them if necessary. Also, flush out the drain line to ensure it is clear. Verify the building piping and flush it out if necessary.
How to fix insufficient water entering the Steris RELIANCE 400 chamber during treatment?
If there is insufficient or no water entering the chamber during treatment, make sure that the supply valves are fully open. The supply line pressure must meet the specifications outlined in the equipment drawing. Open the building and washer supply valves.
Crucial safety information covering personal injury, shock, burn, fire, chemical, and slipping hazards.
Details voltage, amperage, power, noise, environmental conditions, and seismic anchorage.
Steps for running cycles, extending treatments, and managing alarms.
Steps for changing, adding, removing, and creating custom wash cycles.
A guide for the frequency of inspections and maintenance tasks.
Diagnosing and resolving washer malfunctions using troubleshooting charts.