Stille imagiQ2
TM
user manual Emergency handling
n
Emergency handling
Emergency controls
General diagram
Lift the emergency cover off to gain access to emergency controls.
1. Valve for manual Trendelenburg
To be used in case of emergency, if the device is inoperable and the patient needs to be placed in Trendelenburg-
position.
Open the handle for manual longitudinal tilting slowly for a controlled movement.
Close the handle after desired lowering. If the far end of the table is to be raised, a number of people must hold
the end of the table before lengthways tilting is started. Follow instructions on emergency control cover.
2. Extra safety hand control
Use the extra safety hand control in the same way as the ordinary hand control.
Note: Always disconnect damaged hand control before using the safety hand control.