effect. Use this program if you want produce endorphins
at the painful area. The stimulation may reach the
muscular contraction level. If the Pulse Rate will be
increased, the stimulation level must be set to avoid the
muscular contraction.
Mode Constant C Both Pulse Rate and Pulse Width
are adjustable. Default values are PR=70Hz and
PW=50μs characterized by an endorphins stimulating
effect through the Gate-control mechanism. Use this
program in case of ACUTE PAIN. The stimulation level
must be set to avoid the muscular contraction.
Mode MODulation SW (A) Pulse Rate Sweep
modulation. Pulse Rate is automatically modulated within
F1=50Hz and F2=80Hz (adjustable); each frequency step
will last 1 sec. F2 can be changed only if the intensity
level of both channels are set to 0. The Pulse Width is
200μs (adjustable from 50μS to 250μS).
Mode MODulation 2F (B) Pulse Rate Dual
Frequencies modulation. Pulse Rate is automatically
changed every 5 seconds, between F1=70Hz (adjustable)
and F2=5Hz. F1 can be changed only if the intensity level
of both channels are set to 0. The Pulse Width is 50μs
(adjustable from 50μS to 250μS).
Mode Burst A 3 second Burst sequence = 1s on, 2s off.
Both Pulse Rate and Pulse Width are adjustable. Default
values are PR=30Hz and PW=25s. The stimulation
level must be set to produce a gentle muscular
contraction.
Mode Burst B 1.5 second Burst sequence = 0.5s on,
1s off. Both Pulse Rate and Pulse Width are adjustable.
Default values are PR=50Hz and PW=200μs. The
- 19 -