OTHER FEATURES
1. When the unit is turned on, if any of the keys are not
pressed within 5 minutes, it will automatically shut off.
2. When the treatment timer has been set, it begins to
count down one minute by one minute and is
displayed on the LCD. Once it counts down to zero,
the device automatically shuts off.
3. The treatment time will be accumulatively recorded
when the output level is above zero, by pressing
key and “MENU ▼” key for 4 seconds the
treatment time accumulated in minute can be
displayed. Repeat to return back to the previous
normal display in toggle. By pressing key and
“-” key for 4 seconds will clear the treatment time to
zero.
4. When the program changes, the output level will go
down to zero immediately.
5. When the unit is turned on, it will automatically enter
the mode that the unit had worked in before the unit
was turned off.
6. When the battery is low, the battery icon will flash
indicating that the batteries should be replaced.
7. When one or both of the electrodes are not placed
firmly on the skin, the output level goes immediately
to zero.