28
ADDENDUM
Two point thickness calibration
Two point calibration allows a T-Mike E user to bracket a range of thickness by setting the probe
delay on the thinner of two calibration blocks and by setting the velocity on a second, thicker block. In
two point calibration mode the automatic function of the probe zero block on the top of the T-Mike E
is disabled.
The current cal procedure using the probe zero block on the top of the T-Mike E remains the same
and is called one point calibration.
How to select between one and two point calibration
Another option has been added to the gage setup procedure. To enter setup, hold down the [CAL]
key while turning the gage on by pressing and releasing the [ON-OFF] key.