Reprocessed HARMONIC ACE
®
+7, 5mm Diameter Shears with
Advanced Hemostasis Page 8 of 19
Diassessmbly
1. Turn the generator OFF at the power switch.
2. Close the clamp arm and slide the torque wrench over the distal end and up the shaft until the wrench aligns with the flats
on the shaft. Hold by the Hand Piece only, not the instrument handle, and loosen the instrument by turning the wrench
counterclockwise. Continue to loosen by turning the rotation knob manually to completely unscrew the instrument. Do
not untorque the instrument by hand without the torque wrench or damage may occur to the Hand Piece.
Note: Do not use any other means than the torque wrench to detach the instrument from the Hand Piece.
Note: Take care to avoid injury from the blade tip while sliding the torque wrench onto or off of the shaft.
3. Remove the torque wrench by pulling it straight back over the blade.
4. Dispose of the instrument in an appropriate container.
How Supplied
The Reprocessed HARMONIC ACE
®
+7, 5 mm Diameter Shears with Advanced Hemostasis are supplied sterile for single patient
use. Each instrument is shipped with one sterile, single use, gray disposable torque wrench.
Storage and Handling
• Temperature: -20°C to +50°C
• Relative Humidity: 25 to 90%
Compatibility
The Reprocessed HARMONIC ACE
®
+7, 5 mm Diameter Shears with Advanced Hemostasis are designed for use exclusively with
the Generator G11 (GEN11) software version 2013_1, 2014_1, or 2016_1. Software revision can be found under
"System lnformation" in the Generator G11 (GEN11) "Settings" menu. Refer to the Generator G11 (GEN11) Operator's Manual
for more information.
Warranty
Reprocessed Products
Stryker warrants all reprocessed products, subject to the exceptions provided herein, to be free from defects in reprocessing
and to substantially conform to the product specifications contained in the documentation provided by Stryker with the
products for one use in accordance with the instructions for use of such product.
STRYKER SHALL NOT BE LIABLE FOR ANY DAMAGES TO THE EXTENT CAUSED BY ANY DEFECT IN
MATERIAL, WORKMANSHIP OR DESIGN BY THE ORIGINAL MANUFACTURER OF THE PRODUCT OR ANY ACT
OR OMISSION OF THE ORIGINAL MANUFACTURER OF THE PRODUCT.
Products for which Stryker is the Original Manufacturer
Stryker warrants all products for which it is the original manufacturer, subject to the exceptions provided herein, to be free
from defects in design, materials and workmanship and to substantially conform to the product specifications contained in the
documentation provided by Stryker with the products for a period of one year from the date of purchase.
General Warranty Terms Applicable to All Products
TO THE FULLEST EXTENT PERMITTED BY LAW, THE EXPRESS WARRANTY SET FORTH HEREIN IS THE ONLY
WARRANTY APPLICABLE TO THE PRODUCTS AND IS EXPRESSLY IN LIEU OF ANY OTHER WARRANTY BY
STRYKER, EXPRESSED OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OR
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT WILL STRYKER’S LIABILITY
ARISING IN CONNECTION WITH THE SALE OF THE PRODUCT (WHETHER UNDER THE THEORIES OF BREACH
OF CONTRACT, TORT, MISREPRESENTATION, FRAUD, WARRANTY, NEGLIGENCE, STRICT LIABILITY OR ANY
OTHER THEORY OF LAW) EXCEED THE PURCHASE PRICE, CURRENT MARKET VALUE OR RESIDUAL VALUE
OF THE PRODUCTS, WHICHEVER IS LESS. STRYKER SHALL NOT BE LIABLE FOR INDIRECT, SPECIAL,
INCIDENTAL, PUNITIVE OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY BREACH OF WARRANTY OR
UNDER ANY OTHER LEGAL THEORY.