IItteemm IInnssppeecctt
Cot does not lower when you pull the manual backup
release handle with 100 lb (45 kg) or more on the cot
Litter All fasteners are secure
Backrest cylinder operates
Adjust pneumatic cylinder for full range of motion, if
required
Dead stop (6085-001-094) for wear
Base All fasteners are secure
X-frame X-frame expands and retracts
Kickstand (optional) Retracts fully to the transport position
Bolts are tightened
Head section All fasteners are secure
Head section extends and locks
Accessories (optional) All accessories operate (such as IV pole, head extension
and pillow, and restraint extender)
EEvveerryy ssiixx mmoonntthhss oorr 1122 hhoouurrss
Inspect these items every six months or 12 hours, whichever comes first.
IItteemm IInnssppeecctt
Electronic controls Extend cot to raised position, measure and check load
height
Jog function operates
High speed retract works
Switches No damage or wear to either switch
Both switches operate
Litter No bent, broken, or damaged components
No damage or tears on cot grips
Siderails operate and latch
Foot rest operates
Mattress No cracks or tears
Base No bent, broken, or damaged components
Cot retaining post is secure. If not, replace the screw
No excessive damage to X-frame guards
Wheels Free of debris
Head section No bent, broken, or damaged components
EN 48 6086-209-001 Rev F.2

Other manuals for Stryker Performance-PRO XT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Performance-PRO XT and is the answer not in the manual?

Stryker Performance-PRO XT Specifications

General IconGeneral
BrandStryker
ModelPerformance-PRO XT
CategoryMedical Equipment
LanguageEnglish

Related product manuals