Troubleshooting Guidelines (cont’d)
PROBLEM
Sporadic electrical interference is expe-
rienced.
CAUSE
Electrical noise is present.
ACTION
Turn off all electrical equipment not in use
in the operating room.
Relocate electrical equipment; increase
spatial distance.
Plug operating room equipment into
different operating room outlets.
Periodic Maintenance
INTERVAL ACTIVITY
Prior to each use. Inspect, operate and test the handpiece to ensure that it is
working properly. Ensure there are no loose or missing components. Check all
moving parts for free movement. Be alert for unusual sounds or vibrations and note
the operating speed.
Storage and Handling
To ensure the longevity, performance and safety of this equipment, package in original packaging materials when
storing or transporting.