RReeccoommmmeennddeedd sseeppaarraattiioonnss ddiissttaanncceess bbeettwweeeenn ppoorrttaabbllee aanndd mmoobbiillee RRFF ccoommmmuunniiccaattiioonnss eeqquuiippmmeenntt aanndd TTrruuRRiizzee
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
GGuuiiddaannccee aanndd mmaannuuffaaccttuurreerrss ddeeccllaarraattiioonn -- eelleeccttrroommaaggnneettiicc iimmmmuunniittyy
TTrruuRRiizzee is suitable for use in the electromagnetic environment specified below. The customer or the user of TTrruuRRiizzee
should assure that it is used in such an environment.
IImmmmuunniittyy tteesstt
IIEECC 6600660011 tteesstt lleevveell
CCoommpplliiaannccee lleevveell
EElleeccttrroommaaggnneettiicc
eennvviirroonnmmeenntt--gguuiiddaannccee
Electrostatic discharge
(ESD)
IEC 61000-4-2
+8 kV contact
+15 kV air
+8 kV contact
+15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrostatic fast Transient/
burst
IEC 61000-4-4
+2 kV for power supply lines
+1 kV for input/output lines
+2 kV for power supply lines
+1 kV for input/output lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
+1 kV line(s) to line(s)
+2 kV line(s) to earth
+1 kV line(s) to line(s)
+2 kV line(s) to earth
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips, voltage
variations and short
interruptions on power
supply input lines
IEC 61000-4-11
0% U
T
for 0.5 cycles at 0°,
45°, 90°, 135°, 180°, 225°,
270°, and 315°
0% U
T
for 1 cycle
70% U
T
(30% dip in U
T
) for
25 cycles
0% U
T
for 25 cycles
0% U
T
for 0.5 cycles at 0°,
45°, 90°, 135°, 180°, 225°,
270°, and 315°
0% U
T
for 1 cycle
70% U
T
(30% dip in U
T
) for
25 cycles
0% U
T
for 25 cycles
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
TTrruuRRiizzee requires continued
operation during power
main interruptions, it is
recommended that the
device be powered from an
uninterrupted power supply
or a battery.
Power frequency (50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
3333-009-001 Rev H.1 25 EN

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker TruRize 3333 and is the answer not in the manual?

Stryker TruRize 3333 Specifications

General IconGeneral
BrandStryker
ModelTruRize 3333
CategoryMedical Equipment
LanguageEnglish

Related product manuals