Troubleshooting
Event
Code
Description in AccuWin Pro v3 Solution
1
Weakornooscillometricsignal Check position of cuff, tighten cuff.
2
Artifact/ErraticOscillometricSignal Remain still during BP reading.
3
Exceededretrycount
(4 inflate attempts)
Remain still during BP reading.
4
Exceededmeasurementtime
Check air hose connections and
make certain cuff is tight.
85
Reading Aborted
(blocked valves or pneumatics)
Check air hose connections and
make certain air tubing is not
crimped.
86
Reading Aborted (user abort)
PushSTART/STOPbuttontorestart
reading.
87
Reading Aborted
(inflate time-out or air leak)
Check air hose and cuff.
88
Reading Aborted (Safety time-out)
Retryreading,pushSTART/STOP
button. If problems persist, return
unit for servicing.
89
Reading Aborted (cuff over-pressure)
Check for blocked or kinked air
hose.
90
Service Required
(power supply out-of-range or other
hardware problem)
Replace batteries. If problem
persists, return unit for servicing.
91
Service Required
(safety override fitted or autozero
out-of-range)
RetrybypushingSTART/STOP
button. If problems persist, return
unit for servicing.
97
Service Required
(transducer out-of-range)
Return for servicing.
98
Service Required
(A/Dout-of-range)
Return for servicing.
99
Service Required
(EEPROMcalibrationdataCRC
failure)
Unit needs to be recalibrated. Return
for servicing.
In the event you are unable to rectify the error and need assistance, please call our
servicedepartment:intheUS,1.800.421.8626,and919.654.2300,orintheUK
01865.884234.
18