Maintaining and Cleaning the Oscar 2
After use, it is important to perform preventative maintenance to ensure the safe and
efficient operation of the monitor.
Cleaning after use
The Oscar 2unitisnotsterilizable.DONOTimmersethemonitorinanyfluid,or
attempt to clean with any liquid detergents, cleaning agents, or solvents. You may
use a soft, damp cloth to remove dirt and dust from the monitor. If the unit does
becomeimmersedinwater,donotuse;contactourservicedepartment.
You may use a mild disinfectant solution to clean the cuff, strap, belt, and pouch.
Alternatively, you may also wash these items in a washing machine. Remove the
bladder from the Orbitcuffbeforemachinewashing.Washtheseitemsusingwarm
waterandamilddetergent;ifneeded,hangtodry.
Maintenance after use
Visually inspect cables, pneumatic hoses, and the monitor case for cracks, fraying, or
kinks.DONOTusethemonitorifthereareanysignsofdamage.Pleasecontact
our service department.
Maintenance
It is recommended that you check the accuracy of the Oscar 2 once every two
years. If needed, an authorized service center may need to recalibrate the pressure
transducers in the monitor.
19