Release 06/2019
10
Spectroquant
®
NOVA 60A - Operating Manual
To switch on the photometer, open the cover.
The photometer performs a check (Self-Check)
of the entire system and then switches auto-
matically to the concentration measuring mode.
After approx.
5 s:
Commissioning
Self-check of the photometer
Automatic change to the measuring mode,
concentration
Preparing the photometer
The photometer operates at an environmental
temperature of +5 °C to +40 °C. During trans-
port from cold to warm surroundings, conden-
sation can form resulting in the malfunction of
the meter.
Before putting the photometer into service,
wait until it has adapted to the new environ-
mental conditions (see also chapter TECHNICAL
DATA, in the FUNCTIONAL DESCRIPTION docu-
ment [on CD-ROM]).
Place the photometer on a hard, flat surface
and protect it against intensive light and heat.
Line operation
Plug the original power pack into the socket
on the photometer
– Plug the power pack into the line socket
– Switch on the photometer (open the cover).
Battery operation
Charge the battery for approx. 5 hours before
the initial commissioning. To do this:
Plug the original power pack into the socket
on the photometer
Plug the power pack into the line socket and
then the battery will be charged.
During battery operation or when the meter is
at a standstill for longer periods of time, the
battery runs down. This can result in your pho-
tometer no longer being ready for operation.
When the following symbol
is displayed, charge the battery: