Spectroquant
®
NOVA 60A - Operating Manual
Only perform the zero adjustment
against distilled water in an optically
perfect cell.
– Press
In the setup menu, call up the meter setup
submenu.
The following display appears:
After approx.
2 s:
Call up the zero adjustment submenu with
oder
.
– Insert a cell with distilled water.
The message, measuring..., appears on the
display.
Successful zero adjustment for the 10 mm
rectangular cell.
The zero adjustment must be per-
formed separately for each cell type.
Zero adjustment