6
mount camera (not included), unscrew the top ring from the eyepiece adapter. You can
then use it as a 1X C-mount for the camera.
The parfocality between the eyepieces and the trinocular port has been preset before
shipment and should not need adjustment. If necessary, however, you can rotate the
parfocal ring in the eyepiece adapter to adjust the adapter height position and resulting
parfocality.
VI. Caring for your microscope
As with any quality instrument, your microscope should be stored in a cool, dry place.
Use the dust cover to protect the microscope when not in use. If necessary, clean the
lenses with a soft brush or compressed air to gently remove dust.
The specific operation is as follows:
To clean lens surfaces or filters, first remove dust using an air blower. If dust still
persists, use a soft/clean brush or gauze.
A soft gauze or lens tissue lightly moistened with the mixture of alcohol and ether
( ratio : alcohol : 3 and ether: 7) should be used to remove grease or fingerprints.
Use only a mixture of alcohol and ether ( ratio : alcohol : 3 and ether: 7) to remove
immersion oil from objective lenses.
Because the mixture of alcohol and ether ( ratio : alcohol : 3 and ether: 7) is highly
flammable, be careful handling around open flame.
Do not use same area of gauze or lens tissue to wipe more than once.