Page 44 AR-G2/AR 2000ex/AR 1500ex Getting Started Guide
Installation Requirements
It is important to select a location for the instrument using the following guidelines.
Choose a location that is...
In
An indoor area only (A clean environment).
Altitude up to 2000 m.
A temperature-controlled area (5°C to 40°C).
An area where the maximum relative humidity is 80% for temperatures up to 31°C, decreasing linearly
to 50% relative humidity at 40°C.
An area with ample working and ventilation space around the instrument, approximately 2 meters in
length, with sufficient depth for a computer and its keyboard.
On
A stable, vibration-free work surface.
Near
A power outlet. (Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage, instal-
lation category II.)
Your computer for direct connection of serial or ethernet cable or a network port.
Air Bearing Gas Pressure (air or nitrogen) must be clean, dry, oil-free compressed air at 345–690 kPa
gauge (50–100 psig). The dew point should be –20°C or better. Flow rate should be 25 L/min. A 1/4
NPT female connection must be provided for the AR main air supply.
Away from
Dusty environment (pollution degree 2).
Exposure to direct sunlight.
Poorly ventilated areas.
After you have decided on the location for your instrument, refer to the following sections to unpack and
install the AR-G2/AR 2000ex/AR 1500ex Rheometer.