Non-
Volatile
Memory
Information
Stored
User or Case-Spe-
cific Info
Ability to Reset/Clear
Flash #2
TD2u con-
figuration
User-specific
All configuration settings can be
erased and restored to factory
defaults by selecting Setup > Factory
Reset. Configuration settings (exclud-
ing user name and startup mode)
can be reset to defaults using Setup
> Imaging Options >îRestore
Options.
TD2u con-
figuration
Case-specific (if case
ID/notes are enabled)
All configuration settings can be
erased and restored to factory
defaults by selecting Setup > Factory
Reset. Alternately, the Case ID/Notes
can be disabled through Setup >
Case Info Options.
TD2u logs
User-specific and
Case-specific
All logs can be erased by either select-
ing Logs > Erase Logs or Setup > Fact-
ory Reset. Alternately, logs can be
erased individually. When viewing a
specific log entry, select the
[Options] soft key, then select Erase
Log.
Real-Time
Clock
Date/Time None
The Date/Time cannot be restored to
a factory default state. However,
after performing a Factory Reset, the
TD2u prompts you to change the
Date/Time within the Setup Wizard.
The Date/Time can also be changed
by selecting Setup > Date and Time.