EasyManua.ls Logo

Taylor SB25 - Page 17

Taylor SB25
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
Model SB25
Controls and Systems
Calibration
______________________________
The SB25 unit can be calibrated using the supplied
PC software or by using the manual calibration
method through the main keypad.
Target Calibration Settings
S Shaved Ice: 10 oz. (by weight)
S Water Dispense: 4 oz (by volume)
S Cane Dispense: 4 oz (by volume)
PC Calibration
Step 1
Refer to PC Connection on page 12.
Step 2
Select “System Calibration” from the supplied
software. (See Figure 4.)
Step 3
The current system Calibration Factor values for
Shave, Blend, Water and Cane dispense times are
displayed.
Step 4
Use the +/- buttons next to each Calibration Factor
to increase or decrease calibration settings to
achieve desired results. Note that this changes the
Calibration Factor on the shaver/blender system
immediately.
Step 5
Select the Run Cycle button next to each Calibration
Factor to run that particular function with the
Calibration Factor applied.
Step 6
Select “Reset Calibration Factors” to return all
values to 1.0.
Step 7
Select “Main Menu”, select recipe settings, and
select save to unit.
Figure 4

Related product manuals