ASCII protocol format:
ASCII protocol commands:
Baudrate 9600/19200/38400/115200
Data bits 8-bits
Stop bits 1-bit
Parity NONE
Single shot commands
SZ<CR> OK<CR>/ERR<CR> Set a new Zero value
RZ<CR> OK<CR>/ERR<CR> Revert to original calibration Zero
Command Response strings Operation
OP <XXX><CR> OK<CR>/ERR<CR> Open connection to channel XXX
CL<CR> Close channel connection
IV<CR> V:XXXX<CR> Returns Software Version Number (XXXX)
ID<CR> D:DAS<CR> Returns Device Identity (DAS)
IS<CR> S:XXXYYY<CR> Returns Device Status (XXXYYY)
ST<CR> OK<CR>/ERR<CR> Set a new Tare value
RT<CR> OK<CR>/ERR<CR> Revert to Gross mode
GG<CR> G 00000<CR> Get the Gross value
GN<CR> N 00000<CR> Get the Net value
SG<CR> G 00000<CR> Send the Gross value continuously
SN<CR> N 00000<CR> Send the Net value continuously
IM<CR> IM:XXXX<CR> Input Mask - Disable push button functions
IN<CR> IN:XXXX<CR> Read Inputs Status
OM<CR> OM:XXXX<CR> Output Mask - Disable Setpoint outputs
IO<CR> IO:XXXX<CR> Read / Modify Output Status
LI<CR> Lists all menu values
COMMUNICATION PROTOCOLAND COMMANDS
Page 22