Programmer`s manual
EZE30
www.tecsis.de BE 948 a Annex A Page 15
AG Absolute gain calibration. The command returns ERR and has no action unless it is preceded by the CE XXXXX
command, where XXXXX is the current TAC. The gain value (in mV/V, without decimal point) is input as a parameter of AG
followed by the required number of increments as a second parameter, in the format AG 20000 10000. Factory default
setting: 10000 at 2mV/V input.
ZT Zero tracking - this command enables or disables the zero tracking. Parameter = 0 disables the zero tracking and
parameter = 1 enables the zero tracking. Issuing the command without any parameters returns the current ZT value. The
command returns ERR and has no updating action unless it is preceded by the CE XXXXX command, where XXXXX is the
current TAC. Zero tracking will be performed only on results less than +/-0.5d at a rate of 0.4 d/sec where d = display step
size (see DS command). The zero can only be tracked to +/- 20% of max (see CM command). Factory default setting: 0.
FD Factory default settings this command put the EZE30 back to a known state. The data will be written to the
EEPROM and the TAC will be incremented by 1.
NOTE: All calibration and set-up information will be lost by issuing this command.
The command returns ERR and has no updating action unless it is preceded by the CE XXXXX command, where XXXXX is
the current TAC.
CS Save the calibration values - this command results in the calibration values being saved to EEPROM, and causes the
TAC to be incremented by 1. The CS command saves all of the calibration group values, as set by CZ, AZ, CG, AG, CM, CI,
DS, DP and ZT. The command returns ERR and has no updating action unless it is preceded by the CE XXXXX command,
where XXXXX is the current TAC.
3.7 MOTION DETECTION COMMANDS NR, NT

-he signal is steady for the period of time set by
NT, during which it cannot fluctuate by more than NR increments. The stable condition activates the relevant bit of
- .
The func
