QUICK SET UP AND CALIBRATION USING WEIGHTS (Continued)
SELECT THE DECIMAL POINT POSITION ON THE DISPLAY (0, 0.0, 0.00,
0.000 or 0.0000)
Use the UP or DOWN key to set the required decimal point position on the
display. The factory default is no decimal place [0]. The maximum number of
places after the decimal point is four [0.0000]. Please note that this parameter
should be set before the span calibration value [menu 2.1] is entered. Setting
or moving the decimal point position after calibration will not alter the
display resolution
0
T
1
Set the Decimal
Point position
0
T
1
Calibrate the
Zero point
CALIBRATE OR ADJUST THE ZERO POINT (CONVENTIONAL WEIGHING
SYSTEM)
Use this menu point to store a new calibration zero. The display will show the
actual input signal in mV/V. Ensure that the weighing system is at zero or has
no load on it. Pressing the Enter key will store the new zero.
0
T
1
Define the Span
Calibration weight
DEFINE THE SPAN CALIBRATION WEIGHT
This menu point defines the span calibration value. If you intend to calibrate the
span conventionally using weights, you need to enter the value of the span test
weight you intend to apply here. Use the UP/DOWN & MOVE RIGHT keys to
enter the span test weight. For example if you intend to calibrate with 3000 kg
of weight (stepping in 1 kg steps with no decimal places) then enter 3000 here.
0
T
1
Calibrate the Span
CALIBRATE THE SPAN (CONVENTIONAL WEIGHING SYSTEM)
Use this menu point to calibrate the span using weights. Apply the span test
weight [the test weight value is defined in menu 2.1] to the weighing system.
The display will show the actual signal from the weighing system .
Press the Enter key to store the new span value.
in mV/V