EasyManuals Logo

Tektronix 2213 Instruction Manual

Tektronix 2213
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #63 background imageLoading...
Page #63 background image
Performance Check Procedure2213 Service
h. Set the generator output to produce a 1.5-division,
60-MHz display.
i. Repeat part d.
j. Move the generator output from the CH 2 OR Y
input connector to the CH 1 OR X input connector. Set
VERTICAL MODE to CH 1.
k. Repeat part d.
2. Check External Triggering
a. Set:
VOLTS/DIV (both) 10 mV
VERTICAL MODE CH 1
SEC/DIV 20 ns
b. Connect the test setup as shown in Figure 4-1.
c. Set the leveled sine-wave generator to produce a
5-division, 50-kHz display.
d. Set:
VERTICAL MODE CH 2
SEC/DIV 0.2 ms
TRIGGER SOURCE EXT
e. Move the signal from the CH 1 OR X input connector
to the EXT INPUT connector.
f. Set the generator to 2 MHz.
g. CHECKStable display can be obtained by adjusting
the TRIGGER LEVEL control for each switch combination
given in Table 4-6.
h. Remove the 10X attenuator from the test setup and
set the EXT COUPLING switch to DC^-10.
LEVELED
SINE-WAVE
GENERATOR
/
5o-n
CABLE
BNC
T-CONNECTOR
PROBE-TIP-TO-
BNC ADAPTER
10X
PROBE
3827-10
Figure 4-1. Test setup for external trigger and jitter checks.
4-10
REV NOV 1981

Other manuals for Tektronix 2213

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2213 and is the answer not in the manual?

Tektronix 2213 Specifications

General IconGeneral
BrandTektronix
Model2213
CategoryTest Equipment
LanguageEnglish

Related product manuals