k. Reconnect the test setup as shown in Figure 4-1.
l. Set the leveled sine-wave generator to produce a
5-division, 50-kHz display.
m. Set:
CH 2
0.05 fis
On (knob out)
EXT
VERTICAL MODE
SEC/DIV
X I0 Magnifier
SOURCE
Adjustment Procedure—2213 Service
n. Repeat part e.
o. Set the generator to 60 MHz.
p. Repeat parts g and h.
q. Repeat part g.
r. Disconnect the test setup.
REV NOV 1981
5-19