Performance
Check
Procedure
VERTICAL
Equipment
Required
(See
Table
7-1)
Leveled
sine-wave
generator
Calibration
generator
Function
generator
50
1
BNC
coaxial
cable
50
2
precision
BNC
coaxial
cable
50
0
BNC
termination
Dual-input
coupler
Input
COUPLING
Functional
Check
Set:
READOUT
(Intensity)
A
INTEN
For
a
viewable
readout
For
a
viewable
trace
Vertical
MODE
CH
1
and
CH
2
CH
1
and
CH
2
VOLTS/DIV
iv
CH
1
and
CH
2
Input
COUPLING
DG
A/B
SELECT
A
Trigger
Trigger
MODE
AUTO
LEVEL
Trigger
SOURCE
VERT
Trlager
CPLG
Dc
Trigger
SLOPE
_/—
(positive-golng)
Trigger
LEVEL
12
o'clock
Trigger
HOLDOFF
Min
Horizontal
POSITION
12
o'clock
Horizontal
MODE
A
SEC/DIV
0.5
ms
FOCUS
For
best
defined
display
Measurements
All
off
(press
CLEAR
DISPLAY
three
times)
SCOPE
BW
Off
CH
2
INVERT
Off
Set
Vertical
MODE
to
CH
1
(CH
2
off),
Connect
function
generator
(FG
502)
sine-wave
output
to
the
CH
1
Input
via
a
50
1
BNC
coaxial
cable
and
a
50
O
BNC
termination.
7-10