Performance
Check
Procedure
Set:
A
Trigger
CPLG
bc
Horlzontal
MODE
B
A/B
SELECT
B
Trigger
Repeat
100
MHz
NOISE
REJ
Trigger
CPLG
procedure
for
the
B
Trigger.
Single
Sweep
Mode
Set:
Horizontal
MODE
A
A
SEC/DIV
10
ps
A/B
SELECT
A
Trigger
Remove
the
10X
BNC
attenuator
from
the
test
signal
path.
Set
leveled
sine-wave
generator
output
to
produce
a
7.0
division
dis-
play
amplitude
at
50
kHz.
Add
a
10X
and
a
2X
BNC
attenuator
before
the
50
0
BNC
termination.
(Display
should
stably
trigger
with
AUTO
LEVEL
finding
the
correct
trigger
level
setting.)
Set:
A
Trigger
MODE
NORM
CH
1
Input
COUPLING
GND
Trigger
MODE
SGL
SEQ
CHECK—that
the
Trigger
READY
LED
turns
on
and
remains
on.
Set:
A
INTEN
3/4
fully
CW
CH
1
Input
COUPLING
DC
(see
CHECK
below)
CHECK—that
the
TRIG'D
LED
flashes,
and
the
READY
LED
turns
off
after
a
single
sweep
and
readout
display
occurs
when
the
Input
COUPLING
switches
to
DC.