Performance
Check
Procedure
c.
Set
generator
for
2
ms
time
markers.
d.
CHECK—sweep
length
of
the
A
trace
is
greater
than
10
divisions.
6.
Set:
Horizontal
MODE
B
B
SEC/DIV
ims
A/B
SELECT
B
Trigger
Trigger
MODE
RUNS
AFTER
k-
OR
DELAY
Control
cew
to
the
lowest
DELAY
readout
value
B
INTEN
For
a
visible
display
f.
CHECK—the
Delay
Time
readout
is
70.000
ms,
and
the
B
Sweep
length
Is
greater
than
10
divisions.
2.
Horlzontal
POSITION
Range
a.
Set:
Horizontal
MODE
A
Horlzontal
POSITION
Fully
cw
b.
CHECK—that
the
start
of
trace
positions
past
the
center
vertical
graticule
line.
¢.
Set
Horizontal
POSITION
fully
cew.
d.
CHECK—that
the
11th
tlme
marker
Is
positioned
to
the
left
of
the
cen-
ter
vertical
graticule
line.
3.
VAR
SEC/DIV
Range
a.
Set:
SEC/DIV
ims
SEC/DIV
VAR
Fully
cow
Horizontal
POSITION
12
o'clock
b.
Set
time-mark
generator
for
5
ms
time
markers.