ey
Performance
Check
Procedure
CHECK—that
the
sum
of
the
two
readouts
Is
between
999.986
and
1000.014
jis.
Disconnect
the
test
equipment
from
the
2247A.
Rise
Time
Accuracy
Set:
VERTICAL
MODE
CH
1
(CH
2
off)
CH
1
VOLTS/DIV
0.2Vv
A
SEC/DIV
0.2
ps
A
SLOPE
St
Connect
sine-wave
generator
(SG
503)
to
the
CH
1
Input
connector
via
a
50
©
cable
and
a
50
2
terminator.
Set
the
leveled
sine-wave
generator
Frequency
to
1
MHz
and
the
Ampl.
Multiplier
to
X1,
Press
the
COUNTER/TIMER
button.
Select
FREQ.
Adjust
the
sine-wave
generator
for
five
divisions
of
signal.
Adjust
the
sine-wave
generator
for
0.998,000,0
MHz
(998.000.00
kHz)
to
1.002,000,0
MHz
readout
on
the
crt.
Press
the
TIME
button.
Select
(TO
RISE/FALL
MENU),
Select
TRIG
10-90.
Select
RISE.
CHECK—for
a
readout
between
260.11
and
330.23
ns.
NOTE
A
1
MHz
sinewave
has
a
10-90%
rise
time
of
295.167
ns;
20-80%
rise
time
is
204.833
ns.
Move
the
sine-wave
generator
to
the
EXT
COUNTER/TIMER
TIME
BASE
INPUT
on
the
Rear
Panel
of
the
2247A.
Press
the
COUNTER/TIMER
button,
and
select
FREQ.
2247A
Operators
7-59

Related product manuals